Verification of the grid size and angular increment effects in lung stereotactic body radiation therapy using the dynamic conformal arc technique.

DC Field Value Language
dc.contributor.authorPark, HJ-
dc.contributor.authorPark, JY-
dc.contributor.authorSuh, TS-
dc.contributor.authorLee, JW-
dc.contributor.authorKim, MH-
dc.contributor.authorOh, YT-
dc.contributor.authorChun, M-
dc.contributor.authorNoh, OK-
dc.contributor.authorSuh, S-
dc.date.accessioned2014-05-30T06:32:09Z-
dc.date.available2014-05-30T06:32:09Z-
dc.date.issued2013-
dc.identifier.issn0374-4884-
dc.identifier.urihttp://repository.ajou.ac.kr/handle/201003/10301-
dc.description.abstractThe dosimetric effects of variable grid size and angular increment were systematically evaluated in the measured dose distributions of dynamic conformal arc therapy (DCAT) for lung stereotactic body radiation therapy (SBRT). Dose variations with different grid sizes (2, 3, and 4 mm) and angular increments (2, 4, 6, and 10°) for spherical planning target volumes (PTVs) were verified in a thorax phantom by using EBT2 films. Although the doses for identical PTVs were predicted for the different grid sizes, the dose discrepancy was evaluated using one measured dose distribution with the gamma tool because the beam was delivered in the same set-up for DCAT. The dosimetric effect of the angular increment was verified by comparing the measured dose area histograms of organs at risk (OARs) at each angular increment. When the difference in the OAR doses is higher than the uncertainty of the film dosimetry, the error is regarded as the angular increment effect in discretely calculated doses. In the results, even when a 2-mm grid size was used with an elaborate dose calculation, 4-mm grid size led to a higher gamma pass ratio due to underdosage, a steep-dose descent gradient, and lower estimated PTV doses caused by the smoothing effect in the calculated dose distribution. An undulating dose distribution and a difference in the maximum contralateral lung dose of up to 14% were observed in dose calculation using a 10° angular increment. The DCAT can be effectively applied for an approximately spherical PTV in a relatively uniform geometry, which is less affected by inhomogeneous materials and differences in the beam path length.-
dc.language.isoen-
dc.titleVerification of the grid size and angular increment effects in lung stereotactic body radiation therapy using the dynamic conformal arc technique.-
dc.typeArticle-
dc.contributor.affiliatedAuthor박, 혜진-
dc.contributor.affiliatedAuthor김, 미화-
dc.contributor.affiliatedAuthor오, 영택-
dc.contributor.affiliatedAuthor전, 미선-
dc.contributor.affiliatedAuthor노, 오규-
dc.type.localJournal Papers-
dc.citation.titleThe journal of the Korean Physical Society-
dc.citation.volume62-
dc.citation.number11-
dc.citation.date2013-
dc.citation.startPage1672-
dc.citation.endPage1677-
dc.identifier.bibliographicCitationThe journal of the Korean Physical Society, 62(11). : 1672-1677, 2013-
dc.relation.journalidJ003744884-
Appears in Collections:
Journal Papers > School of Medicine / Graduate School of Medicine > Radiation Oncology
Files in This Item:
There are no files associated with this item.

qrcode

해당 아이템을 이메일로 공유하기 원하시면 인증을 거치시기 바랍니다.

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

Browse