Cited 0 times in Scipus Cited Count

Effect of vertebroplasty with bone filler device and comparison with balloon kyphoplasty.

DC Field Value Language
dc.contributor.authorYang, DH-
dc.contributor.authorCho, KH-
dc.contributor.authorChung, YS-
dc.contributor.authorKim, YR-
dc.date.accessioned2016-11-07T10:46:55Z-
dc.date.available2016-11-07T10:46:55Z-
dc.date.issued2014-
dc.identifier.issn0940-6719-
dc.identifier.urihttp://repository.ajou.ac.kr/handle/201003/12781-
dc.description.abstractPURPOSE: To evaluate the effect of vertebroplasty with a bone filler device

compared with balloon kyphoplasty. METHODS: A total of 222 patients underwent

operations from January 2008 to October 2012. One-level fractures numbered 169

(86.7%) cases and two-level fractures numbered 26 (13.3%). A total of 221

vertebral levels were analyzed consequently. Vertebral height, compression ratio,

and segmental Cobb's angle were measured in preoperative and postoperative

lateral X-rays. RESULTS: The compression ratio was the most influential parameter

among three variables. Adjusted postoperative compression ratio was not

significantly different between two operation groups. Bone cement leakage rates

did not differ (p < 0.05). Bone cement distribution was spongy type in the

majority of the vertebroplasty with bone filler device (94.5%), but only in 42.0%

of the kyphoplasty. High bone densitometry readings and long period from

diagnosis to operation were significant risk factors for bone cement leakage.

CONCLUSIONS: Vertebroplasty with a bone filler device could achieve equivalent

compression reduction and bone cement leakage rate, as well as greater

sponge-type bone cement distribution, which were advantages over balloon

kyphoplasty.
-
dc.language.isoen-
dc.subject.MESHBone Cements-
dc.subject.MESHFractures, Compression-
dc.subject.MESHHumans-
dc.subject.MESHKyphoplasty-
dc.subject.MESHMiddle Aged-
dc.subject.MESHRisk Factors-
dc.subject.MESHSpinal Fractures-
dc.subject.MESHTreatment Outcome-
dc.subject.MESHVertebroplasty-
dc.titleEffect of vertebroplasty with bone filler device and comparison with balloon kyphoplasty.-
dc.typeArticle-
dc.identifier.pmid24875381-
dc.identifier.urlhttp://link.springer.com/article/10.1007%2Fs00586-014-3379-7-
dc.contributor.affiliatedAuthor조, 기홍-
dc.type.localJournal Papers-
dc.identifier.doi10.1007/s00586-014-3379-7-
dc.citation.titleEuropean spine journal-
dc.citation.volume23-
dc.citation.number12-
dc.citation.date2014-
dc.citation.startPage2718-
dc.citation.endPage2725-
dc.identifier.bibliographicCitationEuropean spine journal, 23(12). : 2718-2725, 2014-
dc.identifier.eissn1432-0932-
dc.relation.journalidJ009406719-
Appears in Collections:
Journal Papers > School of Medicine / Graduate School of Medicine > Neurosurgery
Files in This Item:
There are no files associated with this item.

qrcode

해당 아이템을 이메일로 공유하기 원하시면 인증을 거치시기 바랍니다.

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

Browse