Influence of electrocautery-induced electromagnetic interference on quantitative electroencephalographic monitoring of hypnosis during general anesthesia: comparison between the ADMS(R) and the BIS VISTATM

DC Field Value Language
dc.contributor.authorKim, DH-
dc.contributor.authorYoo, JY-
dc.contributor.authorKim, JY-
dc.contributor.authorAhn, SH-
dc.contributor.authorKim, S-
dc.contributor.authorMin, SK-
dc.date.accessioned2019-11-13T00:20:37Z-
dc.date.available2019-11-13T00:20:37Z-
dc.date.issued2018-
dc.identifier.issn2005-6419-
dc.identifier.urihttp://repository.ajou.ac.kr/handle/201003/17262-
dc.description.abstractBACKGROUND: Hypnosis monitors analyze small-amplitude electrical signals transmitted from the brain that could be exposed to the electromagnetic field that occurs around the body during electrocautery (ECT). We investigated the influence of ECT on hypnosis monitoring during anesthesia.
METHODS: We simultaneously monitored BIS and uCON during 50 gynecologic oncology surgeries. During the episodes of ECT, we compared the absolute difference (a-Diff) between the baseline index and the most deviated index after ECT over either 30-60 s (ECT30-60) or more than 60 s (ECT > 60) between the monitors. We also investigated the bias and the limits of agreement between the monitors.
RESULTS: Between the two monitors, the a-Diff of ECT30-60 was 1.4 +/- 1.1 for the BIS, which was significantly greater than 0.6 +/- 0.9 for the uCON (P = 0.003), and the a-Diff of ECT > 60 was 16.5 +/- 8.2 for the BIS, which was also significantly greater than 1.4 +/- 1.3 for uCON (P < 0.001). The intra-monitor index differences showed that the BIS during ECT > 60 was significantly greater than that during ECT30-60 (P < 0.001), but the uCON showed no significant difference between ECT30-60 and ECT > 60 (P = 0.056). The estimated bias between the monitors was 6.3 +/- 9.8 and 95% limits agreement was -12.3 to 25.0.
CONCLUSIONS: Prolonged ECT intervention might lead to spurious estimations of quantitative EEG indexes. Therefore, hypnosis should be clinically assessed in combination with scrutinized judgment of relevant clinical symptoms and signs for hypnosis.
-
dc.language.isoen-
dc.titleInfluence of electrocautery-induced electromagnetic interference on quantitative electroencephalographic monitoring of hypnosis during general anesthesia: comparison between the ADMS(R) and the BIS VISTATM-
dc.typeArticle-
dc.identifier.pmid29684988-
dc.identifier.urlhttps://www.ncbi.nlm.nih.gov/pmc/articles/PMC6193592/-
dc.subject.keywordAnesthesia-
dc.subject.keywordElectrocoagulation-
dc.subject.keywordElectroencephalography-
dc.subject.keywordHypnosis-
dc.contributor.affiliatedAuthor김, 대희-
dc.contributor.affiliatedAuthor유, 지영-
dc.contributor.affiliatedAuthor김, 종엽-
dc.contributor.affiliatedAuthor민, 상기-
dc.type.localJournal Papers-
dc.identifier.doi10.4097/kja.d.18.27154-
dc.citation.titleKorean journal of anesthesiology-
dc.citation.volume71-
dc.citation.number5-
dc.citation.date2018-
dc.citation.startPage368-
dc.citation.endPage373-
dc.identifier.bibliographicCitationKorean journal of anesthesiology, 71(5). : 368-373, 2018-
dc.identifier.eissn2005-7563-
dc.relation.journalidJ020056419-
Appears in Collections:
Journal Papers > School of Medicine / Graduate School of Medicine > Anesthesiology & Pain Medicine
Files in This Item:
29684988.pdfDownload

qrcode

해당 아이템을 이메일로 공유하기 원하시면 인증을 거치시기 바랍니다.

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

Browse