dc.description.abstract | BACKGROUND: Band slippage is known to be a troublesome complication of laparoscopic adjustable gastric banding (LAGB), often requiring surgical intervention. To prevent band slippage, a new auxiliary device "S-loop" was developed.
METHODS: From July 2010 to January 2014, a total of 814 LAGBs were performed by a single surgeon. The patients were divided into two groups based on the application of S-loop: conventional LAGB group (n = 378) and S-loop group (n = 436). The operative outcomes were compared between the two groups. RESULTS: The mean operative time and the length of hospital stay were significantly longer in the conventional LAGB group than in the S-loop group (64.3 minutes versus 57.1 minutes: P < .001 and 5.0 hours versus 3.6 hours: P < .001, respectively). The complications occurred in 7.1% of the conventional LAGB group and 1.6% in the S-loop group (P < .001). Slippage was the most common complication: 13 cases were observed in the conventional LAGB group, whereas no slippage was observed in the S-loop group. CONCLUSION: Infra-band fixation using S-loop is a simple and effective method for preventing band slippage compared with the conventional LAGB. | - |
dc.language.iso | en | - |
dc.subject.MESH | Adult | - |
dc.subject.MESH | Female | - |
dc.subject.MESH | Follow-Up Studies | - |
dc.subject.MESH | Gastroplasty | - |
dc.subject.MESH | Humans | - |
dc.subject.MESH | Laparoscopy | - |
dc.subject.MESH | Length of Stay | - |
dc.subject.MESH | Male | - |
dc.subject.MESH | Middle Aged | - |
dc.subject.MESH | Obesity, Morbid | - |
dc.subject.MESH | Operative Time | - |
dc.subject.MESH | Postoperative Complications | - |
dc.subject.MESH | Retrospective Studies | - |
dc.subject.MESH | Weight Loss | - |
dc.title | A Novel Auxiliary Device for Preventing Band Slippage After Laparoscopic Adjustable Gastric Banding: Infra-Band Fixation Using S-Loop | - |
dc.type | Article | - |
dc.identifier.pmid | 29466072 | - |
dc.subject.keyword | bariatric surgery | - |
dc.subject.keyword | laparoscopic adjustable gastric banding | - |
dc.subject.keyword | morbid obesity | - |
dc.subject.keyword | slippage | - |
dc.contributor.affiliatedAuthor | 손, 상용 | - |
dc.type.local | Journal Papers | - |
dc.identifier.doi | 10.1089/lap.2017.0265 | - |
dc.citation.title | Journal of laparoendoscopic & advanced surgical techniques. Part A | - |
dc.citation.volume | 28 | - |
dc.citation.number | 8 | - |
dc.citation.date | 2018 | - |
dc.citation.startPage | 972 | - |
dc.citation.endPage | 976 | - |
dc.identifier.bibliographicCitation | Journal of laparoendoscopic & advanced surgical techniques. Part A, 28(8). : 972-976, 2018 | - |
dc.embargo.liftdate | 9999-12-31 | - |
dc.embargo.terms | 9999-12-31 | - |
dc.identifier.eissn | 1557-9034 | - |
dc.relation.journalid | J010926429 | - |
Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.