Cited 0 times in
Accuracy Analysis of Embletta X100 for the Diagnosis of Obstructive Sleep Apnea and the Assessment of Sleep Structure
DC Field | Value | Language |
---|---|---|
dc.contributor.author | Kim, HJ | - |
dc.contributor.author | Cho, JH | - |
dc.contributor.author | Kang, SY | - |
dc.contributor.author | Choi, BJ | - |
dc.date.accessioned | 2020-01-13T05:03:41Z | - |
dc.date.available | 2020-01-13T05:03:41Z | - |
dc.date.issued | 2018 | - |
dc.identifier.issn | 1229-1498 | - |
dc.identifier.uri | http://repository.ajou.ac.kr/handle/201003/17980 | - |
dc.description.abstract | BACKGROUND AND OBJECTIVES: To measure the accuracy of Embletta X100, a level 2 portable sleep monitoring device, for diagnosis of obstructive sleep apnea and assessment of sleep structure.
MATERIALS AND METHOD: We enrolled 200 consecutive patients who had been referred due to habitual snoring or witnessed apnea during sleep and had undergone standard polysomnography (PSG). We created a simulated situation similar to that of the Embletta X100 using only data from PSG and scored the sleep stage and the apnea-hypopnea index (AHI). Thereafter, the results of PSG and simulated Embletta X100 were compared. RESULTS: Sensitivity, specificity, and positive and negative predictive values of simulated Embletta X100 based on PSG were nearly 100% at three different cutoff values of AHI (5, 15, and 30). Intraclass correlation (ICC) of simulated Embletta X100 based on PSG was also excellent (≥0.9) for most of the sleep-related parameters and respiratory index. However, ICC of sleep stage percent was variable according to sleep stage (>0.9 for N1 and N2, 0.664 for N3, and 0.864 for R). CONCLUSION: Although sleep staging is not very precise, Embletta X100 matches well with PSG overall. | - |
dc.language.iso | en | - |
dc.title | Accuracy Analysis of Embletta X100 for the Diagnosis of Obstructive Sleep Apnea and the Assessment of Sleep Structure | - |
dc.type | Article | - |
dc.subject.keyword | Obstructive sleep apnea | - |
dc.subject.keyword | Portable sleep monitoring | - |
dc.subject.keyword | Embletta X100 | - |
dc.contributor.affiliatedAuthor | 김, 현준 | - |
dc.contributor.affiliatedAuthor | 강, 소영 | - |
dc.contributor.affiliatedAuthor | 최, 병주 | - |
dc.type.local | Journal Papers | - |
dc.identifier.doi | 10.18787/jr.2018.25.1.32 | - |
dc.citation.title | Journal of rhinology | - |
dc.citation.volume | 25 | - |
dc.citation.number | 1 | - |
dc.citation.date | 2018 | - |
dc.citation.startPage | 32 | - |
dc.citation.endPage | 37 | - |
dc.identifier.bibliographicCitation | Journal of rhinology, 25(1). : 32-37, 2018 | - |
dc.embargo.liftdate | 9999-12-31 | - |
dc.embargo.terms | 9999-12-31 | - |
dc.identifier.eissn | 2384-4361 | - |
dc.relation.journalid | J012291498 | - |
Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.