The impact of personal thoracic impedance on electrical cardioversion in patients with atrial arrhythmias

Authors
Roh, SY | Ahn, J | Lee, KN  | Baek, YS | Kim, DH | Lee, DI | Shim, J | Choi, JI | Kim, YH
Citation
Medicina (Kaunas, Lithuania), 57(6). : 618-618, 2021
Journal Title
Medicina (Kaunas, Lithuania)
ISSN
1010-660X1648-9144
Abstract
Background and Objectives—Direct current cardioversion (DCCV) is a safe and useful treatment for atrial tachyarrhythmias. In the past, the energy delivered in DCCV was decided upon empirically, based only on the type of tachyarrhythmia. This conventional method does not consider individual factors and may lead to unnecessary electrical damage. Materials and Methods— We performed DCCV in patients with atrial tachyarrhythmias. The impedance and electrical current at the moment of shock were measured. The human thoracic impedance between both defib-rillator patches and the electric current that was used were measured. Results—A total of 683 DCCVs were performed on 466 atrial tachyarrhythmia patients. The average impedance was 64 ± 11 Ω and the average successful current was 23 ± 6 mA. The magnitude of the electrical current that was successful depended upon the human impedance (linear regression, B = −0.266, p < 0.001) and the left atrial diameter (B = 0.092, p < 0.001). Impedance was directly proportional to body mass index (BMI) (B = 1.598, p < 0.001) and was higher in females than in males (77 ± 15 Ω vs. 63 ± 11 Ω, p < 0.001). Notably, the high-impedance (>70 Ω) group had a higher BMI (27 ± 4 kg/m2 vs. 25 ± 3 kg/m2, p < 0.001) and a higher proportion of females (37% vs. 9%, p < 0.001) than the low-impedance group (<70 Ω). However, thoracic impedance was not an independent predictor for successful DCCV. Con-clusions—Human thoracic impedance was one of the factors that impacted the level of electrical current required for successful DCCV in patients with atrial arrhythmias. In the future, it will be helpful to consider individual predictors, such as BMI and gender, to minimize electrical damage during DCCV.
Keywords

MeSH

DOI
10.3390/medicina57060618
PMID
34199329
Appears in Collections:
Journal Papers > School of Medicine / Graduate School of Medicine > Cardiology
Ajou Authors
이, 광노
Full Text Link
Files in This Item:
34199329.pdfDownload
Export

qrcode

해당 아이템을 이메일로 공유하기 원하시면 인증을 거치시기 바랍니다.

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

Browse