Cited 0 times in
An Evaluation of a New Quantitative Point-of Care Diagnostic to Measure Glucose-6-phosphate Dehydrogenase Activity
DC Field | Value | Language |
---|---|---|
dc.contributor.author | Bahk, YY | - |
dc.contributor.author | Ahn, SK | - |
dc.contributor.author | Jeon, HJ | - |
dc.contributor.author | Na, BK | - |
dc.contributor.author | Lee, SK | - |
dc.contributor.author | Shin, HJ | - |
dc.date.accessioned | 2023-02-21T04:33:35Z | - |
dc.date.available | 2023-02-21T04:33:35Z | - |
dc.date.issued | 2022 | - |
dc.identifier.issn | 0023-4001 | - |
dc.identifier.uri | http://repository.ajou.ac.kr/handle/201003/24666 | - |
dc.description.abstract | Malaria continues to be one of the most crucial infectious burdens in endemic areas worldwide, as well as for travelers visiting malaria transmission regions. It has been reported that 8-aminoquinolines are effective against the Plasmodium species, particularly primaquine, for anti-hypnozoite therapy in P. vivax malaria. However, primaquine causes acute hemolytic anemia in individuals with glucose-6-phosphate dehydrogenase (G6PD) deficiency. Therefore, G6PD deficiency testing should precede hypnozoite elimination with 8-aminoquinoline. Several point-of-care devices have been developed to detect G6PD deficiency. The aim of the present study was to evaluate the performance of a novel, quantitative G6PD diagnostics based on a metagenomic blue fluorescent protein (mBFP). We comparatively evaluated the sensitivity and specificity of the G6PD diagnostic modality with standard methods using 120 human whole blood samples. The G6PD deficiency was spectrophotometrically confirmed. The performance of the G6PD quantitative test kit was compared with that of a licensed control medical device, the G6PD strip. The G6PD quantitative test kit had a sensitivity of 95% (95% confidence interval (CI): 89.3-100%) and a specificity of 100% (95% CI: 94.3-100%). This study shows that the novel diagnostic G6PD quantitative test kit could be a cost-effective and time-efficient, and universally mandated screening tool for G6PD deficiency. | - |
dc.language.iso | en | - |
dc.subject.MESH | Antimalarials | - |
dc.subject.MESH | Glucosephosphate Dehydrogenase | - |
dc.subject.MESH | Glucosephosphate Dehydrogenase Deficiency | - |
dc.subject.MESH | Humans | - |
dc.subject.MESH | Malaria | - |
dc.subject.MESH | Malaria, Vivax | - |
dc.subject.MESH | Point-of-Care Systems | - |
dc.subject.MESH | Primaquine | - |
dc.subject.MESH | Reagent Kits, Diagnostic | - |
dc.title | An Evaluation of a New Quantitative Point-of Care Diagnostic to Measure Glucose-6-phosphate Dehydrogenase Activity | - |
dc.type | Article | - |
dc.identifier.pmid | 36041490 | - |
dc.identifier.url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9441449 | - |
dc.subject.keyword | diagnostics | - |
dc.subject.keyword | G6PD deficiency | - |
dc.subject.keyword | malaria treatment | - |
dc.subject.keyword | mBFP | - |
dc.subject.keyword | primaquine | - |
dc.contributor.affiliatedAuthor | Shin, HJ | - |
dc.type.local | Journal Papers | - |
dc.identifier.doi | 10.3347/kjp.2022.60.4.281 | - |
dc.citation.title | The Korean journal of parasitology | - |
dc.citation.volume | 60 | - |
dc.citation.number | 4 | - |
dc.citation.date | 2022 | - |
dc.citation.startPage | 281 | - |
dc.citation.endPage | 288 | - |
dc.identifier.bibliographicCitation | The Korean journal of parasitology, 60(4). : 281-288, 2022 | - |
dc.identifier.eissn | 1738-0006 | - |
dc.relation.journalid | J000234001 | - |
Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.