Cited 0 times in Scipus Cited Count

Effects of a holmium-166 incorporated covered stent placement in normal canine common bile ducts.

Authors
Won, JH  | Lee, JD | Wang, HJ  | Han, JH  | Kim, JH  | Kim, KH | Park, KB | Itkin, M
Citation
Journal of vascular and interventional radiology : JVIR, 16(5). : 705-711, 2005
Journal Title
Journal of vascular and interventional radiology : JVIR
ISSN
1051-04431535-7732
Abstract
PURPOSE: A specially designed self-expandable covered metallic stent incorporated with beta-emitting radioisotope, Holmium-166 (Ho-166), was developed for delivering intraluminal brachytherapy as well as for internal bile drainage in malignant biliary stricture. The purpose of the study was to demonstrate the safety and tissue response of the radioactive metallic stent on the normal canine common bile duct (CBD) prior to the clinical application.



MATERIALS AND METHODS: Nitinol self-expandable stents (diameter; 4 mm, length; 20 mm) were covered with polyurethane membrane (50 microm thick) containing 21 to 135 muCi of Ho-166 (mean, 77.9 microCi). To prevent migration of stent, the membrane covered only the middle 1 cm of the stent and the ends of the stent were left uncovered. The stents were placed in the CBD of 20 healthy beagle dogs. For control, non-radioactive covered stents were placed in another three dogs. The dogs were killed 3 to 6 months after stent insertion and histopathologic examination of CBD was performed.



RESULTS: There was no stent migration in all cases. Varying degrees of papillary mucosal hyperplasia leading to significant narrowing of the lumen was observed within the lumen of the bare portion and under the mucosal surface of the non-radioactive covered stents. However, fibrosis was noted in Ho-166 coated area, instead of mucosal hyperplasia. Severity of fibrosis was correlated with estimated radiation dose. Despite high dose, there was no perforation of CBD wall. The membrane of Ho-166 coat was disrupted in some cases of 6 months follow-up.



CONCLUSIONS: Holmium-166 incorporated covered stents demonstrated fibrosis of CBD wall and inhibition of ingrowth of mucosal hyperplasia without serious complication such as perforation, while control group showed severe mucosal hyperplasia.
MeSH

DOI
10.1097/01.RVI.0000153113.87723.43
PMID
15872326
Appears in Collections:
Journal Papers > School of Medicine / Graduate School of Medicine > Radiology
Journal Papers > School of Medicine / Graduate School of Medicine > Surgery
Journal Papers > School of Medicine / Graduate School of Medicine > Pathology
Journal Papers > School of Medicine / Graduate School of Medicine > Anesthesiology & Pain Medicine
Ajou Authors
김, 재형  |  왕, 희정  |  원, 제환  |  한, 재호