A comparative study of shade-matching performance using intraoral scanner, spectrophotometer, and visual assessment

Authors
Lee, JH | Kim, HK
Citation
Scientific reports, 14(1). : 23640-23640, 2024
Journal Title
Scientific reports
ISSN
2045-2322
Abstract
This study aimed to explore the clinical applicability of the shade-matching function in intraoral scanners. This study measured the tooth colors of maxillary anterior dentitions of 83 adults using visual matching, a spectrophotometer, and a scanner according to two color systems: VITA Classical (VC) and VITA 3D-Master (V3D). Agreement between each method was assessed by weighted Cohen’s kappa coefficient (KW, α = 0.05). For V3D, the overall agreement between the scanner and spectrophotometer (KW = 0.498) was higher than that between the scanner and visual matching (KW = 0.473). Similarly, the agreement between the scanner and spectrophotometer (KW = 0.283) was higher than that between a scanner and visual matching (KW = 0.140) for VC. Regarding tooth position, the highest agreement between the scanner and spectrophotometer was observed on the right central incisor (KW = 0.542) for V3D. Tooth color measurement with a scanner was comparable to that with a spectrophotometer, especially on the central incisors when using the VITA 3D-Master system. A scanner could serve as an alternative to a spectrophotometer for shade selection. However, color matching should still be visually verified.
Keywords

MeSH

DOI
10.1038/s41598-024-74354-z
PMID
39384854
Appears in Collections:
Journal Papers > School of Medicine / Graduate School of Medicine > Dentistry
Ajou Authors
김, 희경
Full Text Link
Files in This Item:
39384854.pdfDownload
Export

qrcode

해당 아이템을 이메일로 공유하기 원하시면 인증을 거치시기 바랍니다.

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

Browse