Cited 0 times in Scipus Cited Count

Usefulness of the Portable Cystometer

DC Field Value Language
dc.contributor.author윤, 승현-
dc.contributor.author오, 형석-
dc.contributor.author이, 일영-
dc.contributor.author나, 은우-
dc.contributor.author문, 혜원-
dc.contributor.author정, 도영-
dc.date.accessioned2012-01-20T01:44:22Z-
dc.date.available2012-01-20T01:44:22Z-
dc.date.issued2000-
dc.identifier.issn1225-584X-
dc.identifier.urihttp://repository.ajou.ac.kr/handle/201003/5353-
dc.description.abstractObjective: To evaluate the usefulness of a newly devised portable cystometer to measure the pressure of bladder in a low cost and easier accessability.
Method: This study was performed in 47 patients with a conventional instrumental cystometer (Jupiter 8000, Wiest) and the newly devised portable cystometer. In the supine position, we measured the maximal intravesical pressure, abdominal pressure and bladder volume with the portable cystometer just after the measurement of the maximal detrusor pressure and bladder volume with the conventional instrumental cystometer. Paired t-test was utilized to analyze and compare the result.
Results: 1) There was no significant difference in the average maximal detrusor pressures measured by maximal intravesical pressures minus abdominal pressures, which were 38.32⁑20.97 cmH2O by the conventional instrumental cystometer and 40.02⁑20.70 cmH2O by the portable cystometer (p>0.05). 2) There was no significant difference in the average bladder volumes at maximal detrusor pressure, which were 302.13⁑83.92 cc by the conventional instrumental cystometer and 314.04⁑94.17 cc by the portable cystometer (p>0.05).
Conclusion: We conclude that there is no significant difference between the conventional instrumental cystometer and the portable cystometer in the measurement of detrusor pressure and bladder volume. We believe this portable cystometer would be a useful tool to evaluate the function of bladder in a low cost and easier accessability.
en
dc.formatapplication/pdf-
dc.language.isoko-
dc.titleUsefulness of the Portable Cystometer-
dc.title.alternative휴대용 방광 내압 측정기의 유용성-
dc.typeArticle-
dc.subject.keywordCystometer-
dc.subject.keywordUrodynamic study-
dc.subject.keywordDetrusor pressure-
dc.subject.keywordIntravesical pressure-
dc.contributor.affiliatedAuthor윤, 승현-
dc.contributor.affiliatedAuthor이, 일영-
dc.contributor.affiliatedAuthor나, 은우-
dc.type.localJournal Papers-
dc.citation.titleJournal of Korean Academy of Rehabilitation Medicine-
dc.citation.volume24-
dc.citation.number3-
dc.citation.date2000-
dc.citation.startPage463-
dc.citation.endPage468-
dc.identifier.bibliographicCitationJournal of Korean Academy of Rehabilitation Medicine, 24(3). : 463-468, 2000-
dc.relation.journalidJ01225584X-
Appears in Collections:
Journal Papers > School of Medicine / Graduate School of Medicine > Physical Medicine & Rehabilitation
Files in This Item:
Journal of Korean Academy of Rehabilitation Medicine_24(3)_463-468.pdfDownload

qrcode

해당 아이템을 이메일로 공유하기 원하시면 인증을 거치시기 바랍니다.

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

Browse