Cited 0 times in Scipus Cited Count

A Novel Auxiliary Device for Preventing Band Slippage After Laparoscopic Adjustable Gastric Banding: Infra-Band Fixation Using S-Loop

Authors
Son, SY  | Park, YC
Citation
Journal of laparoendoscopic & advanced surgical techniques. Part A, 28(8). : 972-976, 2018
Journal Title
Journal of laparoendoscopic & advanced surgical techniques. Part A
ISSN
1092-64291557-9034
Abstract
BACKGROUND: Band slippage is known to be a troublesome complication of laparoscopic adjustable gastric banding (LAGB), often requiring surgical intervention. To prevent band slippage, a new auxiliary device "S-loop" was developed.
METHODS: From July 2010 to January 2014, a total of 814 LAGBs were performed by a single surgeon. The patients were divided into two groups based on the application of S-loop: conventional LAGB group (n = 378) and S-loop group (n = 436). The operative outcomes were compared between the two groups.
RESULTS: The mean operative time and the length of hospital stay were significantly longer in the conventional LAGB group than in the S-loop group (64.3 minutes versus 57.1 minutes: P < .001 and 5.0 hours versus 3.6 hours: P < .001, respectively). The complications occurred in 7.1% of the conventional LAGB group and 1.6% in the S-loop group (P < .001). Slippage was the most common complication: 13 cases were observed in the conventional LAGB group, whereas no slippage was observed in the S-loop group.
CONCLUSION: Infra-band fixation using S-loop is a simple and effective method for preventing band slippage compared with the conventional LAGB.
Keywords

MeSH

DOI
10.1089/lap.2017.0265
PMID
29466072
Appears in Collections:
Journal Papers > School of Medicine / Graduate School of Medicine > Surgery
Ajou Authors
손, 상용
Files in This Item:
There are no files associated with this item.
Export

qrcode

해당 아이템을 이메일로 공유하기 원하시면 인증을 거치시기 바랍니다.

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

Browse